arrow-backarrow-rightattachmentdownloadenvelopehomeinfo-requestinternetlinklogo-controls-grouplogo-controlsmagnifying-glassmainslider-arrow-leftmainslider-arrow-rightmarkerphoneshare-emailshare-facebookshare-google-plusshare-linkedinshare-pinterestshare-twitterslider-arrow-leftslider-arrow-rightsocial-facebooksocial-linkedinsocial-twittersocial-youtubeswipevideo

Calibration equipment for geotechnical laboratory

  • tecnical specifications
    • Scroll the table
      Force measurement 

      Load cell

      - Linearity: <=0.03%
      - Hysteresis: <=0.03%
      - Repeatability: <=0.01%
       
      Digital readout unit 
      - Accuracy: <=0.02%
      - Standard resolution: (2mV/V) ± 25000 div.
      - Internal resolution: ± 32000 div.
      - Zero and peak functions
      - Auto power off function.
      - Supplied with four 1.5 V AA size alkaline batteries.
      - Typical battery life 1 year.

      Pressure measurement 
      - Linearity and hysteresis: <=0.05%
      - Internal resolution: 65000 div.
      - Zero and peak function
      - Auto power off function.
      - Supplied with four 1.5 V AA size alkaline batteries.

      Displacment Measurement  
      Code 30-WF0652 30-WF0653 30-WF0653/K
      Type Analogue/mechanical Digital Digital
      Maximum travel [mm] 25 50 50
      Resolution [mm] 0.001 0.001 0.001
      Dimesions (wxdxh) [mm] 260x60x100 300x60x107 300x60x107
      Weight [ kg] 0.6 1.5 1.6


  • ordering info
    • LOAD

      Load cell
      30-WF0372/SIT
      5 kN load cell complete with ACCREDIA calibration certificate and stainless steel loading seat.
      30-WF0373/SIT
      50 kN load cell complete with ACCREDIA calibration certificate and stainless steel loading seat.
      NOTE: In order to issue the calibration certificate, the load cell must be ordered complete with digital readout unit

      Digital readout unit
      30-WF6601
      Digital readout unit for load cells
      NOTE: one digital readout unit can be used for both load cells 

      PRESSURE

      30-WF6305/SIT
      50 bar digital pressure gauge complete with ACCREDIA calibration certificate.

      VOLUME CHANGE
      see \#top-loading-balances\#

      DISPLACEMENT 

      30-WF0652
      Calibration device for displacement with analogue micrometer, 25 mm travel x 0.001 mm resolution.
       
      30-WF0652/C
      As above with traceable certificate
       
      30-WF0653
      Calibration device for displacement with digital micrometer, 50 mm travel x 0.001 mm resolution.
       
      30-WF0653/C
      As above with ACCREDIA certificate

      30-WF0653/K
      Calibration device for local strain trasducers and standard linear displacemet trasducers up to 50 mm travel 

      30-WF0653/KC 
      As above with ACCREDIA certificate